Skip to main content

Table 2 Pearson correlation coefficients between corneal modulus and CCT, mean R before and after surgery

From: A novel corneal indentation device for comparison of corneal tangent modulus before and after FS-LASIK in vivo

 

Preoperative modulus

Postoperative modulus

Mean R

 r

0.347**

0.588**

 R2

0.12**

0.35**

CCT

 r

− 0.344**

− 0.554**

 R2

0.12**

0.31**