From: Reliability analysis of hospital infusion pumps: a case study
Days | 142545 | 225973 | 204466 | BO9048 | 143548 | B09320 | 202544 | 142529 | 204553 | BO9070 |
---|---|---|---|---|---|---|---|---|---|---|
0 | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% | 100% |
60 | 81.8% | 81% | 84.1% | 71.8% | 95.6% | 75.7% | 54.8% | 90.3% | 74% | 97.8% |
120 | 69.9% | 70.5% | 73.7% | 56.7% | 89.1% | 59.8% | 44.7% | 79.5% | 63% | 92% |
180 | 60.5% | 62.5% | 65.4% | 45.9% | 82% | 47.9% | 38.40 | 69.2% | 55.2% | 83.5% |
240 | 52.8% | 55.9% | 58.4% | 37.8% | 74.7% | 38.7% | 33.9% | 59.7% | 49.2% | 73.1% |
300 | 46.30% | 50.4% | 52.5% | 31.4% | 67.4% | 31.4% | 30.50 | 51.2% | 44.2% | 61.8% |
360 | 40.8% | 45.7% | 47.3% | 26.3% | 60.4% | 25.7% | 27.7% | 43.6% | 40.1% | 50.6% |
420 | 36.1% | 41.6% | 42.8% | 22.2% | 53.8% | 21% | 25.40 | 37% | 36.6% | 40% |
480 | 32% | 38% | 38.8% | 18.8% | 47.6% | 17.3% | 23:50 | 34.2% | 33.6% | 30.6% |
540 | 28.5% | 34.9% | 35.2% | 16% | 41.9% | 14.3% | 21.80 | 26.3% | 30.9% | 22.7% |
600 | 25.4% | 32% | 32.1% | 13.1% | 36.7% | 11.8% | 20.40 | 22.1% | 28.6% | 16.3% |
660 | 22.6% | 29.5% | 29.2% | 11.8% | 32% | 9.8% | 19.10 | 18.4% | 26.5% | 11.3% |
720 | 20.2% | 27.2% | 20.7% | 10.1% | 27.8% | 8.1% | 17.9% | 15.4% | 24.6% | 7.6% |
780 | 18.1% | 25.2% | 24.4% | 8.8% | 24% | 6.7% | 16.90 | 12.8% | 22.9% | 5% |
840 | 16.2% | 23.3% | 22.4% | 7.6% | 20.7% | 5.6% | 16.00 | 10.6% | 21.4% | 3.2% |
900 | 14.6% | 21.6% | 20.5% | 6.6% | 17.8% | 4.7% | 15.1% | 8.6% | 20% | 1.9% |
960 | 13.1% | 20.1% | 18.8% | 5.7% | 15.2% | 3.9% | 14.4% | 7.3% | 18.7% | 1.1% |
1020 | 11.8% | 18.6% | 17.3% | 5% | 12.9% | 3.3% | 13.60 | 6% | 17.5% | 0.7% |
1080 | 10.6% | 17.3% | 15.9% | 4.3% | 11% | 2.7% | 13.00 | 4.9% | 16.5% | 0.4% |